Analysis of Routine Qa Testing for Conventional Simulators

Shaohua Liang,Xiaowu Deng,Jianyi Chen,Kunyi Zhang
DOI: https://doi.org/10.1007/978-3-540-36841-0_512
2004-01-01
Abstract:[Background and Objective]: Precise simulation is the basis of radiation therapy planning. The geometry accuracy of the simulator, such as the alignment and scales of the field size and angle parameters, especially the l the laser lights would be impacted along with the aging of mechanical parts, changing of running conditions, servicing, etc. A frequent routine quality assurance (QA) testing procedure must be set up and executed. The necessary action criteria of correction can be determined according to the clinical requirement. [Materials and Methods]: Weekly inspection and adjustment of the geometry shifts, including the isocenter precision, field size alignment, collimator and gantry angle scales, coincidence of light and X-ray field, and laser localizations were taken in a whole year, and, correcting adjustments were done after every testing when necessary. All QA parameters were tested for two conventional simulators, the recorded data were analyzed and used to assess and schedule the suitable QA procedure and testing frequency.