Configuration and Calibration of a Flat Field Grating Spectrometer in the Wavelength Range 7–60 Å with a Manson Ultrasoft X-Ray Source

Y. Yang,Z. Shi,Z. Fei,X. Jin,J. Xiao,R. Hutton,Y. Zou
DOI: https://doi.org/10.1088/0031-8949/2011/t144/014064
2011-01-01
Physica Scripta
Abstract:An ultrasoft x-ray and extreme ultraviolet spectrometer built and calibrated in the wavelength range of 7-60 angstrom is reported here. Details of the alignment of this flat field spectrometer with both a laser and a telescope are presented. The light path function rather than a standard calibration function, i.e. a polynomial function, is introduced as the fit function, which gives good agreement with the spectrometer design values and makes the calibration more reliable when extended to the region outside the points used for calibration, compared with a standard calibration function. The calibration results of a Manson ultrasoft x-ray source (model 2) with source targets of Cu, Fe and Ti are presented with all the peaks marked.