Energy Calibration of Transmission Curved Crystal Spectrometer Using Single Filter

Yu Minghai,Hu Guangyue,Yang Tao,An Ning,Wang Qiuping,Zhao Bin,Zheng Jian
DOI: https://doi.org/10.3788/hplpb20142610.102006
2014-01-01
High Power Laser and Particle Beams
Abstract:Theoretical calculation of X-ray three-dimensional diffraction optical path was presented for hard X-ray transmis-sion curved crystal spectrometer.One application of this theory was to study the energy calibration,which was achieved tradition-ally by fitting the K absorption edges of filters or the characteristic peaks of experimental spectra.Obvious error was found in the low energy region of the calibrated spectrum,by taking into account the misalignment of spectrometer or the assembled error of detector.In order to avoid the calibration error and calibrate the spectrometer only by one filter,a new method for energy calibra-tion was presented and verified by experimental measurement of an Ag anode X-ray source.