Focusing Polycapillary To Reduce Parasitic Scattering For Inelastic X-Ray Measurements At High Pressure

paul chow,y m xiao,eric rod,l g bai,g y shen,stanislav v sinogeikin,ningning gao,yongsheng ding,h k mao
DOI: https://doi.org/10.1063/1.4926890
IF: 1.6
2015-01-01
Review of Scientific Instruments
Abstract:The double-differential scattering cross-section for the inelastic scattering of x-ray photons from electrons is typically orders of magnitude smaller than that of elastic scattering. With samples 10-100 mu m size in a diamond anvil cell at high pressure, the inelastic x-ray scattering signals from samples are obscured by scattering from the cell gasket and diamonds. One major experimental challenge is to measure a clean inelastic signal from the sample in a diamond anvil cell. Among the many strategies for doing this, we have used a focusing polycapillary as a post-sample optic, which allows essentially only scattered photons within its input field of view to be refocused and transmitted to the backscattering energy analyzer of the spectrometer. We describe the modified inelastic x-ray spectrometer and its alignment. With a focused incident beam which matches the sample size and the field of view of polycapillary, at relatively large scattering angles, the polycapillary effectively reduces parasitic scattering from the diamond anvil cell gasket and diamonds. Raw data collected from the helium exciton measured by x-ray inelastic scattering at high pressure using the polycapillary method are compared with those using conventional post-sample slit collimation. (C) 2015 AIP Publishing LLC.