A High-Tc Dc-Squid-based Single-Channel MCG and Its Performance
Huiwu Wang,Hongsheng Ding,Lihua Zhang,Xuguang Huang,Weichang Chen,Ziqiang Wang,Ji Feng,Yiping Liu,Shiping Zhao,Yufeng Ren,Hongwei Yu,Fengzhi Xu,Xiaoyin Wang,Genghua Chen,Qiansheng Yang
DOI: https://doi.org/10.1360/02tb9406
2002-01-01
Chinese Science Bulletin
Abstract:A single-channel high-Tc dc-SQUID magnetometer and gradiometer have been developed to record the magnetic field component perpendicular to the human chest generated by heart-beat. Magnetocardiogram (MCG) measurements have been carried out inside a magnetically shielded room. By sequentially adjusting a non-magnetic patient table with 5 cm pitch in X and Y directions, the field signals on a rectangle grid 5 × 5 over the chest area were registered in real time trace point by point with a typical dwell time over 30 cardiocycles each. Utilizing standard electrocardiogram (ECG) recordings as timing reference measured simultaneously with the MCG signals, the MCG data were then averaged and combined to form magnetic field patterns every 10 ms or so. Both the current dipole, which is parallel to MCG measuring plane and produces the vertical magnetic field, and its depth were determined as a function of time in a standard way. We have compared the MCG of healthy hearts with that of a heart with right bundle branch block. Significant differences between their MCGs have been observed.