Flush-Mounted Probe Diagnostics for Argon Glow Discharge Plasma

Liang Xu,Jinxiang Cao,Yu Liu,Jian Wang,Yinchang Du,Zhe Zheng,Xiao Zhang,Pi Wang,Jin Zhang,Xiao Li,Yongqiang Qin,Liang Zhao
DOI: https://doi.org/10.1063/1.4895644
IF: 1.6
2014-01-01
Review of Scientific Instruments
Abstract:A comparison is made between plasma parameters measured by a flush-mounted probe (FP) and a cylindrical probe (CP) in argon glow discharge plasma. Parameters compared include the space potential, the plasma density, and the effective electron temperature. It is found that the ion density determined by the FP agrees well with the electron density determined by the CP in the quasi-neutral plasma to better than 10%. Moreover, the space potential and effective electron temperature calculated from electron energy distribution function measured by the FP is consistent with that measured by the CP over the operated discharge current and pressure ranges. These results present the FP can be used as a reliable diagnostic tool in the stable laboratory plasma and also be anticipated to be applied in other complicated plasmas, such as tokamaks, the region of boundary-layer, and so on.
What problem does this paper attempt to address?