[A Novel Microbeads-Based Chemiluminescence Immunoassay for Detecting Human Thyrotropin].

Yong-ning Wang,Rong-xiu Li,Cheng-wu Li,Yun-hua Gao,Xiao-ying Guo,Zhong-ze Gu
DOI: https://doi.org/10.3321/j.issn:1007-8738.2004.05.034
2004-01-01
Abstract:AIM To establish a low cost and sensitive microbeads-based chemiluminescence immunoassay (CLEIA) for detecting human thyrotropin. METHODS Thyrotropin in sera was captured by an alkaline phosphase-labeled mAb against TSH beta subunit and a FITC-labeled mAb against TSH alpha subunit. Captured thyrotropin was then isolated with immuno-magnetic beads conjugated with anti-FITC antibody and then quantified by chemiluminescence using adamanatane amine as luminescent substrate. RESULTS The sensitivity of this assay is 0.004 mIU/L. The intra-assay CV and the inter-assay CV of the assay were 7.45% and 10.45%, respectively. The recovery rate was 91.4% -102.4%. The examination result of the assay correlates well with that of ACS-180 system. CONCLUSION The CLELA for detecting TSH is low-cost, sensitive, specific and stable, and therefore may have a promising prospect in clinical application.
What problem does this paper attempt to address?