A Highly Sensitive Chemiluminescence Immunoassay Technique for Assaying TSH in Human Serum

QI Su-wen,ZHANG Hui-sheng,LIU Xin-yu,LI Qiao-liang
DOI: https://doi.org/10.3788/fgxb20133408.1089
2013-01-01
Chinese Journal of Luminescence
Abstract:A highly sensitive magnetic particle-based chemiluminescence immunoassay(CLIA) technique for assaying thyroid stimulating hormone(TSH) in serum was described.Fluorescein isothiocyanate(FITC) and N-(aminobutyl)-N-(ethylisoluminol)(ABEI) were used to label two different monoclonal antibodies of anti-TSH.Both of the labeled antibodies combined with TSH to form a sandwiched immunoreaction that was monitored by chemiluminescence(CL) detection.The magnetic particles(MPs) that were coated with anti-FITC antibody served as both the solid phase and the separator.The method showed high sensitivty with a detection limit of 0.01 μIU / mL.The coefficient of variation of intra-assay imprecision and the inter-assay imprecision were less than 10%.The average recoveries were 96% ~ 103%.A correlation analysis against an established automated assay generated a correlation coefficient of 0.993 7.The proposed method demonstrates an acceptable performance for quantifying serum TSH and is suitable for the clinical lab.