Application of europium(III) chelates-bonded silica nanoparticle in time-resolved immunofluorometric detection assay for human thyroid stimulating hormone.

Yulin Zhou,Xiaohu Xia,Ye Xu,Wei Ke,Wei Yang,Qingge Li
DOI: https://doi.org/10.1016/j.aca.2012.01.065
IF: 6.911
2012-01-01
Analytica Chimica Acta
Abstract:Eu(III) chelate-bonded silica nanoparticle was used as a fluorescent label to develop a highly sensitive time-resolved immunofluorometric assay (TrIFA) for human thyroid stimulating hormone (hTSH). The limit of detection of the assay calculated according to the 2SD method was 0.0007mIUL−1 and became 0.003mIUL−1 when serum-based matrix was used for calibrators, indicating that this TrIFA is comparable with the most sensitive assays. The linear range was from 0.005 to 100mIUL−1 of hTSH with coefficient of variation between 1.9% and 8.3%. The correlation study using 204 blood spot samples from newborns showed that the results from this new method were coincident with that of the commercial dissociation-enhanced lanthanide fluorescence immunoassay (DELFIA) system, with a correlation coefficient of 0.938. The fluorescent nanoparticle label allows directly reading the fluorescent signal, omitting the signal development step required for the DELFIA system, and the whole procedure of this assay is fulfilled within 2h. Thus, we developed a novel, sensitive, quantitative and simple nanoparticle label-based TrIFA assay, suitable for routine application in hTSH screening of neonatal hypothyroidism.
What problem does this paper attempt to address?