Chemiluminescent Enzyme Immunoassay for Cardiac Troponin I Detection - Optimization of Experimental Parameters

He Nong-Yue,Guo Hui-Shi,Gu Chun-Rong,Yang Di,Zhang Ji-Nan
DOI: https://doi.org/10.3321/j.issn:0251-0790.2007.02.010
2007-01-01
Abstract:Chemiluminescent immunoassays (CLIA) were intensively studied in antigen/antibody detection, due to the extremely sensitive chemiluminescent detection techniques. Here we present a sensitive immunoassay for cTnI based on this concept. The assay was executed as the traditional ELISA except that two major modifications were made to obtain a rapid test and a high sensitivity. The first modification is that the two-step dual monoclonal antibody "sandwich" principle of traditional ELISA was substituted by a one-step immuno-reaction mode which decreased the turnaround time greatly. The second modification is that a chemiluminescent substrate( AMPPD) and chemiluminescent detection techniques, which elevated the sensitivity remarkably, were used to replace the colorimetric substrate and colorimetric detection method in ELISA. The detection procedure of the assay could be fulfilled within 75 min. A linear range between the luminescent signal current and the concentration of cTnI from 0.04 to 36.2 ng/mL and a detection limit of 0.02 ng/mL were obtained. The established method was tested by determining cTnI in real samples by using ELISA for comparison analysis, and good results were obtained.
What problem does this paper attempt to address?