Electrochemiluminescence Detection of Cardiac Troponin I Based on Au–Ag Alloy Nanourchins

Minghan Wang,Junjie Liu,Xiaoti Qin,Xin Nie,Yifan Dong,Xu Liang,Zhiwei Zhu,Di Yang,Yuanhua Shao
DOI: https://doi.org/10.1039/c9an01904a
2020-01-01
Abstract:In this work, an Au-Ag alloy nanourchin (Au-Ag alloy NU) based electrochemiluminescent (ECL) sensor for the measurement of cardiac troponin I (cTnI) was developed. The as-prepared Au-Ag alloy NUs exhibited higher specific surface area and better conductivity owing to their unique urchin-like morphology, which resulted in excellent electrocatalytic activity towards H2O2 in the luminol-H2O2 ECL system. We have found that the Au-Ag alloy NUs could enhance the ECL signal in the luminol-H2O2 solution. Based on these facts, a facile and label-free ECL immunosensor has been constructed for the analysis of cTnI, a cardiac biomarker, with a wide linear range of 3.5 pg mL-1-350 μg mL-1. This novel ECL immunosensor has good stability and reproducibility, showing potential application in clinical diagnostics. In addition, a non-enzymatic electrochemical sensor for H2O2 was also fabricated, with a wide linear detection range of 100 nM-200 μM, a low limit of detection of 45 nM and a fast response time (less than 2 s).
What problem does this paper attempt to address?