Determination of Iron and Aluminum in Polycrystalline Silicon by Spectrophotometry

HE Fa-lin,CHEN Wen-hui,LUO Xue-tao,CHEN Chao
DOI: https://doi.org/10.3969/j.issn.1000-7571.2010.10.014
2010-01-01
Abstract:The polycrystalline silicon(silica powder) was double dissolved with hydrofluoric acid and nitric acid.The interference of fluorinion with aluminum was eliminated by double smoking of perchloric acid.The great error caused by low sensitivity of iron was solved by using 3 cm cuvette.The content of iron and aluminum in polycrystalline silicon were accurately determined by spectrophotometry.The effect of coloring conditions on coloring reaction was investigated to obtain the optimal experimental conditions.The results showed that in HAc-NaAc buffer solution at pH 3-5,Fe2+ could form a stable red complex with 5 mL of 1,10-phenanthroline(0.25 g/L) after 10 min and Beer's law was obeyed for iron in the range of 2-10 μg/50 mL.In hexamethylenetetramine buffer solution at pH 5.5-6.1,the Al3+ could form a stable purple-red complex with 5 mL of chromeazurol-S(0.3 g/L) after 15 min,and Beer's law was obeyed for aluminum in the range of 4-16 μg/50 mL.The content of iron and aluminum in polycrystalline silicon(silica powder) with purity of 99.99% could be accurately determined by spectrophotometry.The determination results were compared with those obtained by ICP-AES.The relative error was about 5 %.
What problem does this paper attempt to address?