ICP-AES Method with Multicomponent Spectral Fitting Tests Silicon Element in High Silicious Content Alloys

He Yong
DOI: https://doi.org/10.3969/j.issn.1007-1865.2010.07.049
2010-01-01
Abstract:A method for determination of silicon element in high silicious content alloys by ICP-AES combined with mixed acids digestion was developed in the paper.And the silicious precipitation during the digestion process was directly sloved by add hydrofluoric acid.Cooling the digested solution sufficient to avoid the silicon tetrafluoride evaporate.Multicomponent spectral fitting technical was used to eliminate the matrix interference.The digested solution cold directly be used to test other elements of the alloy.