Study on ICP-AES Determination Method for 18 Trace Elements in Highly Calcareous Sediments from the South China Sea

Qiang Sun
2007-01-01
Abstract:Calcium and magnesium often interfere the measurements of other elements in using inductive coupling plasm atomic emission spectrometer(ICP-AES) to analyze the trace elements in highly calcareous sediments from the South China Sea.In this paper,a series of procedures including optimization of instrument operating parameters,deduction of background,correction of spectral line interference and matching of datum are proposed for eliminating the element interference and the datum effects of calcium and magnesium and raising the sensitivity and precision of this method.An ICP-AES method suitable for determining the trace element contents in highly calcareous marine sediments is established by making a great deal of experiments on sediment samples from the South China Sea and verifying the experimental results with state standard samples.The detection limit and relative standard deviation(RSD) of the method are 1.55 to 271 μg/L and 4.54%(n=6),respectively.