In-vitro test apparatus and experimental study of mechanics properties of coronary stents

程洁,周啸,李俐军,倪中华
DOI: https://doi.org/10.3969/j.issn.1001-0505.2010.02.024
2010-01-01
Abstract:In order to improve the design and manufacturing of coronary stents,on machine vision technique was developed.Relevant experiments were designed and performed to evaluate an independent designed thin-walled cutting tubular 316L stent with its transportation performance,radial strength during the intervention,and hemodynamics in pulsatile flow after the implantation.It is found that the stent has good transportation performance with radial strength of a maximum value of 180 kPa.It is also proved that stenting induces the reduction of the local wall shear stress,especially in pre-stent and post-stent regions,which may be a risk factor for restenosis by neointimal hyperplasia.The results show that the appartus can meet the requirement of the multiple mechanics-properties tests of stents.