Preparement of Percutaneous Valved Stent in Vitro

周永新,邵杰,孙林,李刚,梅运清,王永武
2008-01-01
Abstract:Objective To prepare percutaneous valved stent and discuss its design,performance and the feasibility of trans-catheter implantation.Methods Self-expandable nitinol stents were made by 0.2 mm NiTi shape memory alloy(15~19 mm in diameter) and its physical performances including component,shaperestoring temperature,magnetism,tensile force,extending ratio,super-elasticity,durability,corruption speed and surface trait etc were examined.Bovine pericardium was cross-linked then be shaped and be sutured into stents.Fluid passing test,pre-expanding and releasing test and static test of pressure in tube were performed in all valved stents.Results The nitinol self-expanding stents had excellent physical properties in performance.Their shape-restoring temperature ranged from 30℃to 40℃,and they had no-magnetism. Their tensile force,extending ratio,super-elasticity,durability,corruption speed,surface trait all reached related standards.in vitro test the closure of the valved stent leaflets were effective,and fluid flows were not restricted in the opposite direction.The valved stents could be released through catheter,then expanded completely and could fix fast in the tube.Conclusion The percutaneous valved stent which is made of nitinol self-expanding stent and bovine pericardium has excellent physical performance and the function of valve is satisfactory.It is well designed and can be implanted through catheter.
What problem does this paper attempt to address?