Long-term safety and absorption assessment of a novel bioresorbable nitrided iron scaffold in porcine coronary artery
Jian-Feng Zheng,Zi-Wei Xi,Yang Li,Jia-Nan Li,Hong Qiu,Xiao-Ying Hu,Tong Luo,Chao Wu,Xin Wang,Lai-Feng Song,Li Li,Hai-Ping Qi,Gui Zhang,Li Qin,Wan-Qian Zhang,Xiao-Li Shi,Shu-Han Wang,De-Yuan Zhang,Bo Xu,Run-Lin Gao
DOI: https://doi.org/10.1016/j.bioactmat.2022.01.005
IF: 18.9
2022-11-01
Bioactive Materials
Abstract:This study aimed to investigate the long-term biocompatibility, safety, and degradation of the ultrathin nitrided iron bioresorbable scaffold (BRS) in vivo, encompassing the whole process of bioresorption in porcine coronary arteries. Fifty-two nitrided iron scaffolds (strut thickness of 70 μm) and 28 Vision Co-Cr stents were randomly implanted into coronary arteries of healthy mini-swine. The efficacy and safety of the nitrided iron scaffold were comparable with those of the Vision stentwithin 52 weeks after implantation. In addition, the long-term biocompatibility, safety, and bioresorption of the nitrided iron scaffold were evaluated by coronary angiography, optical coherence tomography, micro-computed tomography, scanning electron microscopy, energy dispersive spectrometry and histopathological evaluations at 4, 12, 26, 52 weeks and even at 7 years after implantation. In particular, a large number of struts were almost completely absorbed in situ at 7 years follow-up, which were first illustrated in this study. The lymphatic drainage pathway might serve as the potential clearance way of iron and its corrosion products.
engineering, biomedical,materials science, biomaterials