Finite Element Analysis of Expansion Behavior of Coronary Stents with Different Materials

ZHANG Qing-bao,WANG Wei-qiang,QI Min,LIANG Dong-ke,YANG Da-zhi
DOI: https://doi.org/10.3321/j.issn:1001-9731.2007.01.039
2007-01-01
Journal of Functional Biomaterials
Abstract:Coronary stent is an important device used in percutaneous transluminal coronary angioplasty(PTCA),which is usually implanted to support the stenosed artery.Its expansion behavior depending on inflating of balloon and characters of recoil happening while the balloon is retrieved are the important aspects which should be concerned in operation.In this paper,the finite element method(FEM) was applied to systematically analyze the dilated mechanical behavior of 316L stainless steel and L605 stents,which is influenced by width and thickness of strut,the dilated scale of stents,vessel and stent materials used.The results show that the dilated mechanical behavior of stents depends greatly on the material used.The internal pressure needed for L605 stent to expand and recoil of L605 stent is larger than that of 316L stainless steel.And increasing strut width or thickness can improve the critical internal pressure needed for stent to expand fast.And the change of stent length while expanding is only associated with its structure and final dilatation state.In conclusion,FEM can quantify some mechanical behavior of stent and help designers to optimize stent system.
What problem does this paper attempt to address?