Evaluation of the room shielding thickness of Hi-Art tomotherapy system

Haikuan LIU,Jinhai WU,Naigu GU,Yiming GAO,Li WANG,Weiqin HUANG,Fengxian WANG
2010-01-01
Abstract:In this paper, we calculate and evaluate the room shielding thickness of a Hi-Art tomotherapy system, which is a new type of radiotherapy facility. Due to the self-shielding of the accelerator, only scattered beam and beam leakage were considered in calculating the room shielding thickness. The radiation field of the tomotherapy system was used as the basic data to calculate the shielding thickness of every 15° solid angle. The maximum shielding thickness required of each shielding wall was at the position with the angle of 15°, and the calculated shielding thickness were 1023, 975, 917, 1460, 1147 and 1189 mm for the east wall, south wall, west wall, north wall, the roof and the floor, respectively. According to the calculation results, all shielding walls, ceiling and floor could meet the requirement of the radiation protection, but the north wall thickness of 1200 mm was a little thinner.