Higher Order Harmonics Contribution and Suppression in Metrology Beamline

Zhou Hongjun,Wang Guanjun,Zheng Jinjin,Hou Tonglin,Qiu Keqiang
DOI: https://doi.org/10.1117/12.885285
2010-01-01
Abstract:Using the house-made transmission grating (3300 l/mm) and IRDAXUV100G (USA) photodiode detector, higher order harmonics contributions have been measured in the spectral radiation standard and metrology beamline at NSRL and the contributions of the different orders have been analyzed. In order to accurately calibrate the performance of optical elements, higher order harmonics must be efficiently suppressed. Selecting the suitable filter in different wavelength, the higher order contributions can be efficiently suppressed. In wavelength region between 5 nm and 12 nm with proper Zr filters, the contributions of higher order intensity are less than 1.25%, after modified by quantum efficiency of the detector, the higher order contributions are restricted to less than 0.62 %. In wavelength region between 13 and 17 nm with Si filter, the higher order contributions are almost zero. In wavelength region between 18 nm and 34 nm with proper Al filter, the contributions of higher order intensity are less than 8.06%, after modified by quantum efficiency of the detector, the higher order contributions are restricted to less than 3.08 %. In wavelength region between 35 and 40nm with Al/Mg/Al filter, the higher order contributions are restricted to less than 10.00 % after modified by quantum efficiency of the detector.