A Gas Filter System of U14C Beamline at National Synchrotron Radiation Laboratory

ZHANG Tai-chang,ZHU Ai-guo,HONG Xin,PAN Yang,SHAN Xiao-bin,SHENG Liu-si,ZHANG Yun-wu,QI Fei
DOI: https://doi.org/10.3969/j.issn.0253-2778.2007.04.045
2007-01-01
Abstract:A gas filter system was designed for a newly built U14C beamline at National Synchrotron Radiation Laboratory to eliminate high order harmonic radiation and maintain other systems of the beamline in ultrahigh vacuum environment.The differential pumping system contains a gas cell and three differential pumping regions.Testing results indicate that the beamline pressure is still kept at lower than 1.09×10-6 Pa when the pressure of Ar in the gas cell is lower than 1.333 kPa,that the vacuum of the beamline is around 2.53×10-7 Pa with 1.333 kPa Ar in the gas filter,and that high order harmonic radiation could be suppressed better than 99.97%.