[Trace Detection of Ammonia at 1.531 Microm].

Jia Hui,Guo Xiao-yong,Cai Ting dong,Zhao Wei-xiong,Wang Lei,Tan Tu,Zhang Wei-jun,Gao Xiao-ming
DOI: https://doi.org/10.3964/j.issn.1000-0593(2009)12-3173-04
2009-01-01
Spectroscopy and spectral analysis
Abstract:A compact instrument based on the off-axis integrated-cavity output spectroscopy (ICOS) technology was developed for sensitive measurements of gas mixing ratios (ammonia in air) at room temperature by using fiber-coupled distributed feedback (DFB) diode laser operating at 1.531 microm. The absorption line of ammonia at 6 528.764 cm(-1) was chosen for trace detection. The mirrors' effective reflectivity R2 of 0.996 9 was first calibrated by carbon dioxide under this condition, and the cavity 35.8 cm in length as an absorption cell could yield an optical path of presumably 115.46 m. As a result, a minimum detectable concentration of approximately 2.66 ppmv (S/N-3) at the total pressure of 100 torr was obtained. Then the lock-in amplifier was added in the system to acquire the second harmonic signal by combination of wavelength modulation technology, which could better suppress background noise and improve the signal-to-noise ratio, and a detection limit of 0.293 ppmv (S/N-3) was achieved eventually. This work demonstrated the potential of the system for a range of atmospheric species sensing in the future.
What problem does this paper attempt to address?