NO Concentration Sensing at 1.79 Μm Transition Using Tunable Diode Laser Absorption Spectroscopy

Lifang Zhang,Fei Wang,Jian-Hua Yan,Ke-Fa Cen
DOI: https://doi.org/10.1063/1.4872112
2014-01-01
AIP Conference Proceedings
Abstract:A tunable diode laser-based sensor for measuring the concentrations of NO has been developed. In this paper, the measured spectra of NO 3v band (the R branch), H2O and CO2 lines in the spectral region from 5565 to 5588 cm at 600K were analyzed and compared with each other to select optimum transitions for measuring NO concentration. The absorption transitions R (7.5) were selected according to HITRAN database to minimize interference from water vapor and CO2. The line strength is about 10(-23) CM-1 / (molecule/cm(2)), and a minimum detectable of 30 ppm was obtained with wavelength modulation spectroscopy (WMS-2f) method. The 1.79 mu m distributed feed-back (DFB) diode laser has a narrow line width. The measurement was validated in a gas cell at 300K, where NO concentrations using both scanned-wavelength direct absorption (DA) and wavelength modulation spectroscopy (WMS-2f) strategies were validated against calibrated mixture concentrations. Then the wavelength modulation measurement was validated in heated gas cell in a tubular furnace (373 similar to 573K). The laboratory use LabVIEW program to realize data acquisition and to realize the real-time display of the concentration of the gas through direct absorption method. In the laboratory, we also analyze the concentrations of the NO using WMS at different temperature (300K-573K).