Electrical Characterization of Flexible Neural Microelectrode Array for Retinal Prosthesis

Wu Yibo,Xu Ailan,Hui Chun,Ren Qiushi,Li Gang
DOI: https://doi.org/10.19650/j.cnki.cjsi.2008.12.012
2008-01-01
Abstract:Substrate-integrated microelectrode arrays have become one of the hottest topics in the world, which are used in implantable prosthesis for recording of neural signals or electrical stimulation of nerves. A new type of flexible microelectrode arrays based on parylene is developed in this paper. And the electrical performance of the flexible neural MEAs is characterized using electrochemical methods. Test results suggest that the impedance value of the parylene-based MEAs is in the range of 2000-100Ω, and its phase angle is -20° in medium frequency band. With these electrical performances, functional electrical stimulation can be realized easily for implantable system.
What problem does this paper attempt to address?