Fabrication of Carbon Film-Based Flexible Neural Microelectrode Array
ZHU Zhuang-hui,ZHOU Liang,ZHANG Hua,LI Gang,JIN Qing-hui,ZHAO Jian-long
DOI: https://doi.org/10.3969/j.issn.1672-6030.2012.02.016
2012-01-01
Nanotechnology and Precision Engineering
Abstract:To achieve long-term safe and effective stimulation and to avoid erosion of electrodes,a novel flexible retinal stimulation microelectrode array was developed using pyrolyzed photoresist carbon film as a conducting material in this paper.Firstly,the patterned photoresists were pyrolyzed to produce conducting carbon films at high temperature.Then,a photosensitive polyimide(Durimide 7510) was used as the substrate material.By patterning with photolithography techniques,sequentially transferring the carbon film,bonding two insulting layers and etching the sacrifice layer,the flexible carbon film microelectrode arrays were successfully fabricated.Experimental results show that the charge storage capacity of the carbon film microelectrodes is 6.625 mC/cm2,which is about 12.4 times greater than that of the gold microelectrodes,indicating that the carbon film microelectrodes have a good electrochemical stability and a large charge injection capability,which is beneficial to their long-term implantation and safe work.This flexible carbon film microelectrode array can be a promising candidate for retinal prosthesis.