Functional Abstract and design of bioelectrical impedance measurement system

Wang Chao,Zhang Qiang,Huang Chunyan,Wang Huaxiang
DOI: https://doi.org/10.19650/j.cnki.cjsi.2008.01.008
2008-01-01
Abstract:In order to implement universality, interoperability and transplantability in bioelectrical impedance measurement system, using the idea of abstract, this paper divides the system into four parts: signal source module, excitation selection module, measurement selection module and demodulation module. In the signal source module, excitation selection module and measurement selection module, MCUs (Micro Controller Unit) are used as the main control unit. The universal interface circuit and communication protocol are designed, which implements the uniformity of the data exchange. The multiplication demodulation circuit, which can demodulate the mixed frequency signal, is designed in the demodulation module. Finally, a mixed frequency bioelectricar impedance measurement system was constructed using these abstract modules.
What problem does this paper attempt to address?