Design of Demodulation Circuit for Multi-Frequency-One-Time Bioelectrical Impedance Measurement System

Zhao Jinchuang,Li Zhigang,Qiu Bin,Fu Wenli,Huang Chong
DOI: https://doi.org/10.16526/j.cnki.11-4762/tp.2012.09.019
2012-01-01
Abstract:To acquire the impedance of different tissues and organs at the same time,this paper designs a demodulation circuit for the Multi-Frequency-One-Time(MFOT) bioelectrical impedance measurement system based on frequency selection and analog multiplier module.The demodulation system works at 10Hz~1MHz.Its common mode rejection ratio(CMRR) is 80dB at 500KHz.The demodulation results with the error less than 1.92% reflect the real part and imaging part parameters of impedance models clearly.This design satisfies the requirement of bioelectrical impedance measurement system with high CMRR and excellent measurement accuracy.