Mixing frequency based bioelectrical impedance measurement

Wang Chao,Qian Xiangchen,Wang Xiangyu,Wang Huaxiang,Yan Yong
DOI: https://doi.org/10.3321/j.issn:0254-3087.2007.06.001
2007-01-01
Abstract:In order to acquire the bioelectrical impedance information at different frequencies concurrently, a new measurement system was developed using a mixing frequency approach. The system design is based on the virtual reference vector method and multiplication demodulation technique. The implementation of the system hardware is achieved using a modular structure. A digital signal processor controls various modules in the system, and the system software realizes configuration of the digital signal processor, serial communication and data presentation, etc. A reference impedance analyzer (Agilent 4294A) is used to provide reference data in order to evaluate the developed system. Results obtained on a human body show that the system performs well in terms of functionality and accuracy.
What problem does this paper attempt to address?