Design of A Multi-Parameter Bioimpedance Measurement System

Fanchang Meng,Shuxin Chen,Jiali Wang,Feng Xu,Yuguo Chen,Ke Li
DOI: https://doi.org/10.1109/icarm62033.2024.10715997
2024-01-01
Abstract:Bioimpedance provides vital information for a variety of clinical application. Most current bioimpedance device are confined in their large volume, high cost and single detectable parameter. This study aimed to design a novel four-channel, multi-parameter bioimpedance acquisition device with lower volume and wearable design. This new device was designed based on an AD5933 impedance analyzer chip, enabling real-time acquisition of bio-impedance values, bio-impedance phase angle, and bio-impedance ratio through a small extension circuit for impedance acquisition and an analog switching circuit based on ADG888. Results showed that the sampling rate of the novel device could reach up to 100 Hz with a resolution of $0.01 \Omega$ and errors within ± 1%. Application of this novel device in bioelectrical defibrillation showed that the impedance information could successfully guide the onset and energy outputs of the defibrillation. This device may provide a novel avenue for wearable bioimpedance measurement and have a potential application in bioelectrical defibrillation.