A High-Speed Bioelectrical Impedance Spectroscopy System Based on the Digital Auto-Balancing Bridge Method

Nan Li,Hui Xu,Wei Wang,Zhou,Guofeng Qiao,David D-U Li
DOI: https://doi.org/10.1088/0957-0233/24/6/065701
IF: 2.398
2013-01-01
Measurement Science and Technology
Abstract:A novel bioelectrical impedance spectroscopy system based on the digital auto-balancing bridge method improved from the conventional analogue auto-balancing method is presented for bioelectrical impedance measurements. The hardware of the proposed system consists of a reference source, a null detector, a variable source, a field programmable gate array, a clock generator, a flash and a USB controller. Software implemented in the field programmable gate array includes three major blocks: clock management, peripheral control and digital signal processing. The principle and realization of the least-mean-squares-based digital auto-balancing algorithm is introduced in detail. The performances of our system were examined by comparing with a commercial impedance analyzer. The results reveal that the proposed system has high speed (less than 3.5 ms per measurement) and high accuracy in the frequency range of 1 kHz-10 MHz. Compared with the commercial instrument based on the traditional analogue auto-balancing method, our system shows advantages in measurement speed, compactness and flexibility, making it suitable for various bioelectrical impedance measurement applications.
What problem does this paper attempt to address?