Calibration Methods of Irradiance Responsivity of UVA Detectors and International Comparison

DAI Cai-hong,YU Jia-lin,YU Jing,LIN Yan-dong,XIONG Li-min,YIN Chun-yong
2008-01-01
Abstract:Five methods for measuring the irradiance responsivity of UVA detectors are discussed and analyzed in detail,including absolute spectral irradiance method based on standard lamps,absolute spectral power responsivity of detectors,absolute thermoelectric radiometer method,electrically calibrated pyroelectric radiometer method and method of comparing with the NIM standard UVA radiometer.On the basis of the above theories and methods,NIM took part in the APMP international comparison of Irradiance Responsivity of UVA Detectors PR-S1 in 2004.The comparison results showed that the consistency between the NIM value and the international reference one is the best among the seven participants.The deviation of NIM′s value from the international reference one is only-0.57%(k=2) for UV365 irradiance responsivity,and-0.53%(k=2) for UVA irradiance responsivity.In addition,the uncertainty of broadband ultraviolet irradiance under the defined conditions was cut down to 2.0%(k=1) from 10%(k=1).