Keitz Method for Measuring of 254 Nm Radiant Flux in Low Pressure Mercury Lamps

ZHANG Yuan,SUN Ming-hua,ZHU Shao-long,ZHANG Shan-duan
DOI: https://doi.org/10.15943/j.cnki.fdxb-jns.2010.02.020
2010-01-01
Abstract:Two methods of measuring UV radiant flux are introduced.The Keitz formula is used to calculate the 254 nm radiant flux of low pressure mercury lamp (LPM).The difference between linear lamp approximation, cylinder lamp approximation and Keitz formulas is discussed.The results show that the Keitz method and gonioradiometry result in the error less than 1 % for the distance between UV sensor and LPM 2.5-4 times of the lamp length.The radiant efficiency of 254 nm line is about 40% for LPM with outer diameter 16 mm,length 0.8 m, and power 120 W.