Automatic Measurement for Light Distribution Curve of Ultra High Pressure Mercury Lamps

YU Zhi-ping,ZHENG Zhen-rong,LU Wei
DOI: https://doi.org/10.3785/j.issn.1008-973x.2005.08.037
2005-01-01
Abstract:Through analyzing the effects of ultra high pressure (UHP) mercury lamp's dimension and its distance from the detector on the measurement of light distribution curve, an automatic equipment based on computer interface technique and photoelectric conversion technique was constructed to obtain the exact light distribution curve of UHP lamps. Program time delay and software filtering were used to delete the interference of the drive current and the mechanical vibration, so the measuring precision was improved. The data measured by the equipment not only can comply with the requirement of illumination system simulation, but also can help the light source designer to ameliorate the UHP lamp's burner and electrode.