Radiometric calibration model of uncooled infrared focal plane array

Xue-jin SUN,Jian LIU,Shi-jun ZHAO,Dong-li ZHAI,Jie-tai MAO
DOI: https://doi.org/10.3969/j.issn.1009-3443.2008.04.017
2008-01-01
Abstract:In order to measure radiance with Uncooled InfraRed Focal Plane Array (UIRFPA) under different conditions, a calibration experiment method based on the standard radiant source was designed, and a radiometric calibration model which involved detector temperature effect was proposed according to the calibration test data and theoretical analysis. The experiment results show that the UIRFPA has good response linearity, wide dynamic range, that the detector temperature effect is linear, and that the retrieval standard deviation of the radiometric calibration model is 0.72 w/(m2·sr).