Minipermanent Magnet High-Current Microwave Ion Source

Zhizhong Song,Shixiang Peng,Jinxiang Yu,Jianchuan Ming,Zhongxi Yuan,Feng Qian,Zhiyu Guo
DOI: https://doi.org/10.1063/1.2150807
IF: 1.6
2006-01-01
Review of Scientific Instruments
Abstract:A 2.45GHz high-current microwave ion source with permanent magnet was developed at Peking University. The source body is about 10cm in diameter and 10cm in height, and its weight is less than 5kg. It can operate in both pulse and continuous wave (cw) modes. In pulse mode, more than 100mA hydrogen ion current was extracted from a 5mm ion extraction aperture, when the microwave power is 500W and the extraction voltage is 45kV, which corresponds to a beam density of 500mA∕cm2. In cw mode, up to 60mA of hydrogen ion current can be extracted, and the beam density is 300mA∕cm2 at the same conditions. The proton ratio is about 80% and the normalized root-mean-square emittance is less than 0.1πmmmrad in both modes.
What problem does this paper attempt to address?