Accurate assessment of f′1 value of fine photometer heads

Jiangen Pan,Haiping Shen,Huajun Feng,Guofei Xiang
DOI: https://doi.org/10.1088/0026-1394/42/4/007
2005-01-01
Metrologia
Abstract:For an accurate assessment of the f(1)(') value of a photometer head it is important to determine the maximum step size for sampling and an appropriate bandwidth of the monochromator for an acceptable uncertainty contribution. A simulation is carried out to evaluate these uncertainty contributions for fine photometer heads, with a mismatch characteristic f(1)(') <= 3% and uncertainty contributions Delta f(1)(') <= 0.1% taken as the limits for acceptance. Assuming a stable measurement set-up, the results show that a 5 nm step size and a 5 nm bandwidth or less are needed for this assessment. The uncertainty of the f(1)(') value caused by the wavelength uncertainty of the monochromator is analysed additionally. For extended bandwidths a method is proposed to correct the spectral responsivity data of the photometer head and reduce the bandpass uncertainty contribution effectively.