Effect of Unequal Countercurrent Blood Vessels on the Temperature Distribution Due to HIFU

陈琦,白景峰,陈亚珠
DOI: https://doi.org/10.3321/j.issn:1006-2467.2004.01.031
2004-01-01
Abstract:The paper investigated the effect of unequal countercurrent blood vessels on the temperature distribution during phased high intensity focused ultrasound (HIFU) treatment. A bioheat transfer model based on three heat balance equations was adapted for 3-D transient simulations. The simulation of this model using finite element method was carried out for the tissue with the 50~300 μm diameter paired blood vessels. Contribution from the vessel diameter, blood velocity and blood perfusion rate were compared and the influence on 3-D temperature distribution was estimated. The results show that the paired vessel diameter and the blood velocity are the crucial factors determining the distribution of temperature for HIFU treatment. Moreover, the model is able to give more precise information about the lesion volume of heated tissue.
What problem does this paper attempt to address?