Relative impedance variation based temperature monitoring and dose control for high intensity focused ultrasound therapy

Qingyu Ma,Gepu Guo,Juan Tu,Dong Zhang,Shaotong Feng
DOI: https://doi.org/10.1121/1.4970749
2016-01-01
Abstract:High intensity focused ultrasound (HIFU) can be used as a non-invasive and efficient therapy approach to treat tumors by heat coagulation with less metastasis, although it is still limited by inaccurate temperature monitoring and efficacy evaluation. Based on the relationship between temperature and electrical impedance, a model of impedance measurement during HIFU therapy was established. Numerical studies on the distributions of acoustic pressure, temperature and conductivity were simulated, and the impedances in the model were calculated with the electrical potential and current density. It is proved that, for a fixed acoustic power in the focal region, the relative impedance variation (RIV) linearly increases with the increasing treatment time and is inversely proportional to acoustic power, while the relative impedance variation rate (RIVR) shows a linear relationship with the acoustic power. The treatment time that is required for the temperature in the focal region to reach 70 degree is inversely p...