In vivo hyperthermia effect induced by high-intensity pulsed ultrasound

Weicheng Cui,Juan Tu,Jooha Hwang,li qian,Tingbo Fan,Dong Zhang,Jinghai Chen,Weizhong Chen
DOI: https://doi.org/10.1088/1674-1056/21/7/074301
2012-01-01
Chinese Physics B
Abstract:Hyperthermia effects (39-44 degrees C) induced by pulsed high-intensity focused ultrasound (HIFU) have been regarded as a promising therapeutic tool for boosting immune responses or enhancing drug delivery into a solid tumor. However, previous studies also reported that the cell death occurs when cells are maintained at 43 degrees C for more than 20 minutes. The aim of this study is to investigate thermal responses inside in vivo rabbit auricular veins exposed to pulsed HIFU (1.17 MHz, 5300 W/cm(2), with relatively low-duty ratios (0.2%-4.3%). The results show that: (1) with constant pulse repetition frequency (PRF) (e.g., 1 Hz), the thermal responses inside the vessel will increase with the increasing duty ratio; (2) a temperature elevation to 43 degrees C can be identified at the duty ratio of 4.3%; (3) with constant duty ratios, the change of PRF will not significantly affect the temperature measurement in the vessel; (4) as the duty ratios lower than 4.3%, the presence of microbubbles will not significantly enhance the thermal responses in the vessel, but will facilitate HIFU-induced inertial cavitation events.
What problem does this paper attempt to address?