Controllable in Vivo Hyperthermia Effect Induced by Pulsed High Intensity Focused Ultrasound with Low Duty Cycles

Juan Tu,Joo Ha Hwang,Tao Chen,Tingbo Fan,Xiasheng Guo,Lawrence A. Crum,Dong Zhang
DOI: https://doi.org/10.1063/1.4754113
IF: 4
2012-01-01
Applied Physics Letters
Abstract:High intensity focused ultrasound (HIFU)-induced hyperthermia is a promising tool for cancer therapy. Three-dimensional nonlinear acoustic-bioheat transfer-blood flow-coupling model simulations and in vivo thermocouple measurements were performed to study hyperthermia effects in rabbit auricular vein exposed to pulsed HIFU (pHIFU) at varied duty cycles (DCs). pHIFU-induced temperature elevations are shown to increase with increasing DC. A critical DC of 6.9% is estimated for temperature at distal vessel wall exceeding 44 degrees C, although different tissue depths and inclusions could affect the DC threshold. The results demonstrate clinic potentials of achieving controllable hyperthermia by adjusting pHIFU DCs, while minimizing perivascular thermal injury. (C) 2012 American Institute of Physics. [http://dx.doi.org/10.1063/1.4754113]
What problem does this paper attempt to address?