An Experimental Study on Optimal Temperature Latitude and Duration Within the Focal Region by High Intensity Focused Ultrasound

熊六林,石军,郑珠颖,薛卫成,黄晓波
DOI: https://doi.org/10.3969/j.issn.1672-6448.2009.01.004
2009-01-01
Abstract:Objective To evaluate the damage degree of pig's liver and pancreas at 45℃-60℃ within different durations in vivo,then identify the optimal temperature latitude and duration within the focal region target by high intensity focused ultrasound.Methods The liver and pancreas of pigs were exposed in vivo to defined temperatures of 45±1℃、50±1℃、55±1℃、60±1℃ for varying durations of 15,30,60,90 or 120 seconds using a radio-frequency ablation probe.Exposed tissue was obtained immediately following treatment or on day 3 following exposure.Histopathologic evaluation was performed using light microscopy,transmission electron microscopy.Results Pancreas and liver exposed to 45±1℃ for 15-60 seconds,demonstrated no discernable changes immediately following treatment by light microscopy.Transmission electron microscopy performed on tissue exposed for 60 seconds showed cell organ swelling without rupture.At 50±1℃ for 15 seconds,light microscopy immediately showed obvious cell swelling and electron microscopy showed breakdown of mitochondria membrane and endocytoplasmic reticulum.On day 3 there was evidence of coagulative necrosis by light microscopy.Tissue exposed to 55-60℃ for 15 seconds demonstrated coagulative necrosis by light microscopy performed on tissue obtained immediately following treatment.Conclusions The temperature of 55-60℃ for 15 seconds may be the optimal temperature and duration within the focal region of HIFU.