In Vitro Peripheral Irradiation of Ox-liver Tissue Using High Intensity Focused Ultrasound

YIN Li,ZOU Jian-zhong,WU Feng,PU Rui,WANG Qi,LIU Fang
DOI: https://doi.org/10.3969/j.issn.0258-4646.2012.11.007
2012-01-01
Abstract:Objective To investigate the efficiency of peripheral irradiating on ablation ox-liver tissue using high intensity focused ultrasound(HIFU).Methods A Total of 30 cases of ox-liver tissues were randomly divided into 2 groups:the peripheral irradiation group received peripheral irradiation in region-of-interest but not in central zone;the traditional irradiation group was irradiated completely from deep area to superficial area.Results There was no significant difference of the volume of coagulation necrosis between the two groups(P 0.05),while the exposure time,energy and energy efficiency factor(EEF) of the peripheral irradiation group were less than the traditional irradiation group(P 0.05).Conclusion The coagulation necrosis of peripheral irradiation was the same as the traditional irradiation,while the exposure time and energy of were significantly reduced.The peripheral irradiation was more efficient than the traditional irradiation on HIFU ablation.