The Impact of Large Blood Vessel Location on Thermal Dose Distribution During Focused Ultrasound Surgery for Hepatic Tumor Ablation

ZHANG Chenxi,ZHANG Su,ZHANG Zhao,CHEN Yazhu
DOI: https://doi.org/10.3969/j.issn.1005-0930.2006.03.007
2006-01-01
Abstract:The effects of presence of a large blood vessel and its distance to a tumor on lesion sizes during focused ultrasound surgery were studied.A finite element method(FEM) was used to obtain the temperature distribution for bioheat transfer equation(BHTE) with the consideration of the large blood vessels convection effect.In the 3-D FEM simulation,a 4-mm diameter,10-mm height cylindrical tumor tissue was heated using different heating schemes.The distance between the blood vessel and the tumor tissue varied from 1mm to 3mm.The model without blood vessel was also studied.Simulation results quantitatively show that the domain of thermal lesion can effectively cover the desired therapeutic region with short ultrasound exposure times(2s),however, if the large blood vessel is near the tumor with the distance of 1mm,the blood flow may play an important role in the thermal dose distribution on tumor boundary.Underdosed areas might be present from which the tumor can regenerate.When the distance of the longer blood vessel from the tumor is longer than 3mm,the cooling effect of large blood vessels is not significant in focused ultrasound surgery.