A Sequential Injection On-Line Column Preconcentration System for Electrothermal AAS Determination of Thallium in Geochemical Samples

ZR Xu,SK Xu,ZL Fang
2020-01-01
Atomic Spectroscopy
Abstract:A sequential injection system for on-line sorbent extraction preconcentration in electrothermal atomic absorption spectroscopy was developed for the determination of trace thallium in geochemical samples. The TIBr 4 -1 complex was adsorbed on a 20-μL micro-column (located at the tip of the furnace sampling probe) packed with XAD-8 resin. After sequentially aspirating separate zones of acetone, rinsing acid, and sample (pretreated with bromine) into a 2.5-m long, 1-mm i.d. holding coil, the flow was reversed and directed to the column. Sample loading, analyte adsorption, column rinsing and analyte elution were achieved within a single reversed syringe stroke. The adsorbed analyte was eluted into the furnace with 50 μL acetone. Mutual mixing between sample, rinsing acid, and eluent were prevented by separating the zones with small air segments during metering. Tightening of column packing was avoided by a slight back-suction through the column after each operational cycle. With 1-mL sample loading, an enrichment factor of 15 was obtained with a detection limit of 18 ng/L thallium (3σ). A precision of 2.4% RSD (n=11, 4 μg/L) and a sampling frequency of 11/hour were achieved. The method was applied to the analysis of geochemical samples. The results were in good agreement with the certified values of standard reference geochemical materials.