Solution Cathode Glow Discharge – Atomic Emission Spectrometry (SCGD-AES) With an Interference Filter for Spectral Discrimination for the Determination of Thallium in Surface Water

Jinmei Wang,Guoxing Fei,Peichao Zheng,Nantian Yan,Guanghui Chen,Wei Li,Daming Dong,Lianbo Guo
DOI: https://doi.org/10.1080/00032719.2024.2333341
2024-03-31
Analytical Letters
Abstract:Solution cathode glow discharge atomic emission spectrometry (SCGD-AES) using wide-bandpass (10-nm) interference filters (IF) as a spectral discrimination device was used for the determination of thallium (Tl) in surface water. The use of wide-bandpass interference filters greatly reduces the cost of traditional detection instruments. The influence of solution flow rate, discharge current and pH on the detection limit was characterized. Moreover, matrix effect experiments were performed to study the effect of different concentrations of Na + , K + , Ca 2+ , and Mg 2+ on the Tl signal intensity. Using a solution flow rate of 2.8 mL/min, a discharge current of 50 mA, and pH of 1.0 with HNO 3 as the supporting electrolyte, SCGD coupled with an interference filter provided a detection limit of 7.76 μg/L for Tl in water samples, comparable to traditional SCGD-AES. Furthermore, the spiked recovery rate of Tl is quantitative in real water samples and comparable to the performance of inductive coupled plasma – optical emission spectrometry (ICP-OES).
chemistry, analytical
What problem does this paper attempt to address?