Preliminary research on dose response of LiF thermoluminescence dosimeter to irradiation of 49.3MeV/u carbon ion beam

Qiang Li,Guiling Wang,Zengquan Wei,Wenjian Li,Bingrong Dang,Jufang Wang,Guangming Zhou
1999-01-01
Abstract:LiF(Mg,Ti) thermoluminescence dosimeter (TLD) was applied to measure dose response to irradiation of 49.3MeV/u 12C ion beam supplied by HIRFL. Moderate variance of dose-response efficiency was found when particle fluence increases from l× 107 to 2×108/cm2. The average dose-response efficiency 57% was obtained. This phenomenon implies that LiF can be used to measure absorbed dose in a certain range of particle fluence. Moreover, the result is in a good agreement with the calculated dose-response efficiency 54% to 50MeV/u 12C ion beam, which was predicted by the model of heavy-ion track structure at GSI, Germany.