TL Response of LiF:Mg,Cu,P (GR200A and GR207A) Exposed to High-Energy 12C Ions

X. Y. Zhang,L. Chen,P. Luo,Y. Q. Ju,H. B. Zhang,F. Ma,H. L. Ge,Y. B. Zhang,B. Wan,Y. Y. Li,X. W. Xu,W. Mao
DOI: https://doi.org/10.1016/j.radmeas.2014.10.004
IF: 1.743
2014-01-01
Radiation Measurements
Abstract:The characteristics of (LiF)-Li-nat.,7:Mg,Cu,P thermoluminescent detectors exposed to high-energy C-12 (165 MeV/nucleon) ions and Co-66 gamma-rays have been investigated. Through a careful analysis of the glow curves by the GlowFit program, it is found that the relative amplitude of satellite peaks 3, 5, 6 and 7 to peak 4, measured after irradiated with high-energy C-12 ions, is higher than that with Co-60 gamma-rays for both types of TL, detectors. It is observed that "(LiF)-Li-nat.,7:Mg,Cu,P thermoluminescent detectors would also exhibit a good linear dose response when the delivered dose of high-energy C-12 ions is in the range from 100 mGy to 1600 mGy. Furthermore, the relative efficiency of the (LiF)-Li-nat.,7:Mg,Cu,P thermoluminescent detectors is independent of the absorbed dose in the linear dose response range. Compared to previously published data, the relative efficiency of the (LiF)-Li-7:Mg,Cu,P TL detectors decreases with increasing LET values. (C) 2014 Elsevier Ltd. All rights reserved.