Preliminary study on characteristics of in vivo dosimetry using TLD during tumor radiotherapy

WANG Yu-long,FENG Yan,LI Long-gen,XIA Jun-ding
DOI: https://doi.org/10.3969/j.issn.0253-9926.2013.07.009
2013-01-01
Abstract:Objective To study the elementary properties of thermo luminescent dosimeter(TLD) and the elementary function of read system for its clinical use.To provide experimental basis for developing its clinical use of in vivo measurements within dose of clinical radiotherapy.Methods Using TLD(51 tablets) and ionization chamber(NE-Farmer 2581-A TYPE) as well as build-up cap(to consist of LTS-TYPE plastics) under clinical conditions of external irradiation with the AECL-780C 60Co,the characteristics of thermo luminescent dosimeter were investigated in phantom and were verified with ionization chamber in the same phantom.Results The surface of TLD was clean and smooth.Its diameter,thickness and weight per square centimeter were 3.00 mm,0.80 mm and(29.4 ±0.9)mg/cm2,respectively.Within 1 000 cGy,the relationship between thermo luminescence values and doses of irradiation was linear,and the correlativity was close(r=0.99),with a sensitivity of 1.00±0.09.The deviation of 0.66% was derived from cGy rating of repeated irradiation.Conclusion The properties for TLD can be achieved by a good choice,and present excellent reproducibility and stability,negligible dependence on temperature,linear response versus dose from some cGy to some Gy,independence of dose-rate and high spatial resolution.It entirely accord with the norm of ICRU.As shown in this part,TLD is the well-established method used for in vivo dosimetry.