Thermoluminescent dosimeters (TLD-100) for absorbed dose measurements in alpha-emitting radionuclides

Andrew J. White,Sean P. Jollota,Cliff G. Hammer,Ahtesham U. Khan,Larry A. DeWerd,Wesley S. Culberson
DOI: https://doi.org/10.2139/ssrn.4645566
IF: 1.787
2024-03-30
Applied Radiation and Isotopes
Abstract:Early works that used thermoluminescent dosimeters (TLDs) to measure absorbed dose from alpha particles reported relatively high variation (10%) between TLDs, which is undesirable for modern dosimetry applications. This work outlines a method to increase precision for absorbed dose measured using TLDs with alpha-emitting radionuclides by applying an alpha-specific chip factor (CF) that individually characterizes the TLD sensitivity to alpha particles. Variation between TLDs was reduced from 21.8% to 6.7% for the standard TLD chips and 7.9%–3.3% for the thin TLD chips. It has been demonstrated by this work that TLD-100 can be calibrated to precisely measure the absorbed dose to water from alpha-emitting radionuclides.
radiology, nuclear medicine & medical imaging,nuclear science & technology,chemistry, inorganic & nuclear
What problem does this paper attempt to address?