Two Types of Ion Gun Equipped Sputter-Initiated Resonant Ionization and Time-Of-Flight Mass Spectroscopy and Its Application in Microanalysis of Minerals

L Jin,ST Dai,GX Dong,CM Li,SL Wang,JH Tian,DY Chen
DOI: https://doi.org/10.1063/1.1148789
IF: 1.6
1998-01-01
Review of Scientific Instruments
Abstract:An apparatus, based on sputter-initiated resonant ionization spectroscopy and combined with time-of-flight mass spectroscopy, has been recently developed for microanalysis of mineral samples and other solid materials. This equipment has two kinds of ion gun, which make it possible to do more convenient and accurate microanalysis of complicated mineral samples. With this apparatus, detection and size measurement of fine gold grains trapped in pyrite at low concentrations have been carried out. The detection limit of this apparatus for gold is 73 ppb. The apparatus also has the function of secondary electron imaging. This gives very helpful information for geologists to evaluate the inborn state of gold in mineral samples and to improve the extraction rate of gold from minerals.
What problem does this paper attempt to address?